A超角膜测厚仪和Pentacam系统测量散瞳前后角膜中央厚度的比较Comparison of Central Corneal Thickness Before and After Mydriasis Measured with A-ultrasound Pachymetry and Pentacam Scheimpflug System
金玉梅;艾凤荣;罗岩;李莹;
JIN Yu-mei;AI Feng-rong;LUO Yan;LI Ying;Department of Ophthalmology,Peking Union Medical College Hospital,Chinese Academy of Medical Sciences &Peking Union Medical College;
摘要(Abstract):
目的比较应用A超角膜测厚仪和Pentacam三维眼前节分析系统测量的散瞳前后角膜中央厚度的差异。方法2011年8月至11月北京协和医院成人屈光不正患者65例(130眼),均先后应用Pentacam三维眼前节分析系统和A超角膜测厚仪进行散瞳前后角膜中央厚度的测量,散瞳剂采用复方托吡卡胺滴眼液。结果 A超角膜测厚仪测量的散瞳前后角膜中央厚度值分别为(536.70±39.60)和(548.31±39.22)μm,散瞳后的测量值明显高于散瞳前(P=0.000);Pentacam三维眼前节分析系统测量的散瞳前后角膜中央厚度值分别为(539.57±39.73)和(545.99±39.47)μm,散瞳后的测量值亦明显高于散瞳前(P=0.000)。两种方法散瞳前角膜中央厚度的测量值比较,Pentacam三维眼前节分析系统测量值明显高于A超角膜测厚仪(P=0.000);而两种方法散瞳后的测量值比较,差异无统计学意义(P=0.081)。结论散瞳后角膜中央厚度测量值较散瞳前略有增加,故散瞳前测量值更能准确、真实地反映角膜中央厚度,有助于为角膜屈光手术方式的选择和手术安全性评价提供参考。
Objective To compare the difference of central corneal thickness(CCT) before and after mydriasis measured with A-ultrasound pachymetry and Pentacam three-dimensional anterior segment analysis system.Methods From August 2011 to November 2011,the CCT of 130 eyes from 65 adult patients with myopia in Peking Union Medical College Hospital were measured with A-ultrasound pachymetry and Pentacam Scheimpflug system before and after mydriasis using compound tropicamide eye drops. Results The CCT values obtained with A-ultrasound pachymetry before and after mydriasis were(536. 70±39. 60) μm and(548. 31±39. 22) μm,respectively. The value after mydriasis was significantly higher than that before mydriasis(P=0. 000). The CCT values obtained with Pentacam three-dimensional anterior segment analysis system before and after mydriasis were(539. 57±39. 73) μm and(545. 99±39. 47) μm,respectively. The value after mydriasis was also significantly higher than that before mydriasis(P=0. 000). CCT value obtained with Pentacam three-dimensional anterior segment analysis system was significantly higher than A-ultrasound pachymetry before mydriasis(P=0. 000);however,there was no significant difference in CCT between the two methods after mydriasis(P=0. 081). Conclusions Values of CCT measured after mydriasis are higher than before mydriasis. Therefore,the values of CCT measured before mydriasis may be more accurate,which can help to provide a better option for choosing operative methods and evaluating the safety of refractive surgery.
关键词(KeyWords):
角膜中央厚度;Pentacam三维眼前节分析系统;A超角膜测厚仪;散瞳
central corneal thickness;Pentacam three-dimensional anterior segment analysis system;A-ultrasound pachymetry;mydrasis
基金项目(Foundation):
作者(Author):
金玉梅;艾凤荣;罗岩;李莹;
JIN Yu-mei;AI Feng-rong;LUO Yan;LI Ying;Department of Ophthalmology,Peking Union Medical College Hospital,Chinese Academy of Medical Sciences &Peking Union Medical College;
Email:
DOI:
参考文献(References):
- [1]Wolfs RC,Klaver CC,Vingerling JR,et al.Distribution of central corneal thickness and its association with intraocular pressure:The Rotterdam Study[J].Am J Ophthalmol,1997,123:767-772.
- [2]Realini T,Lovelace K.Measuring central corneal thickness with ultrasound pachymetry[J].Optom Vis Sci,2003,80:437-439.
- [3]Salouti R,Nowroozzadeh MH,Zamani M,et al.Comparison of anterior chamber depth measurements using Calilei,HR Pentacam,and OrbscanⅡ[J].Optometry,2010,81:35-39.
- [4]Piňero DP,AlióJL,Alesón A,et al.Corneal volume,pachymetry,and correlation of anterior and posterior corneal shape in subclinical and different stages of clinical keratoconus[J].J Cataract Refract Surg,2010,36:814-825.
- [5]Buehl W,Stojanac D,Sacu S,et al.Comparison of three methods of measuring corneal thickness and anterior chamber depth[J].Am J Ophthalmol,2006,141:7-12.
- [6]Atchison DA.Comparison of peripheral refractions determined by different instruments[J].Optom Vis Sci,2003,80:655-660.
- [7]Bourges JL,Alfonsi N,LalibertéJF,et al.Average 3-dimensional models for the comparison of OrbscanⅡand Pentacam pachymetry maps in normal corneas[J].Ophthalmology,2009,116:2064-2071.
- [8]Lattinore MR Jr,Kaupp S,Schallhorn S,et al.Orbscan pachymetry:implications of a repeated measures and diurnal variation analysis[J].Ophthalmology,1999,106:977-981.
- [9]叶娅,沈政伟,尹禾,等.Pentacam系统测量正常人眼表面麻醉前后中央角膜厚度的对比研究[J].眼科新进展,2011,1:68-70.
- [10]Rosa N,Lanza M,Borrelli M,et al.Comparison of central corneal thickness measured with Orbscan and Pentacam[J].J Refract Surg,2007,23:895-899.
- [11]De Sanctis U,Missolungi A,Mutani B,et al.Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry[J].Am J Ophthalmol,2007,144:712-718.
- [12]Ciolino JB,Khachikian SS,Belin MW.Comparison of corneal thickness measurements by ultrasound and Scheimpflug photography in eyes that have undergone laser in situ keratomileusis[J].Am J Ophthalmol,2008,145:75-80.
- 角膜中央厚度
- Pentacam三维眼前节分析系统
- A超角膜测厚仪
- 散瞳
central corneal thickness - Pentacam three-dimensional anterior segment analysis system
- A-ultrasound pachymetry
- mydrasis
- 金玉梅
- 艾凤荣
- 罗岩
- 李莹
JIN Yu-mei- AI Feng-rong
- LUO Yan
- LI Ying
- Department of Ophthalmology
- Peking Union Medical College Hospital
- Chinese Academy of Medical Sciences &Peking Union Medical College
- 金玉梅
- 艾凤荣
- 罗岩
- 李莹
JIN Yu-mei- AI Feng-rong
- LUO Yan
- LI Ying
- Department of Ophthalmology
- Peking Union Medical College Hospital
- Chinese Academy of Medical Sciences &Peking Union Medical College